Hemodynamic study of unenhanced magnetic resonance angiography using spatial labeling with multiple inversion pulses sequence: a phantom study

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

单位: [1]Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Dept Radiol, Wuhan, Peoples R China
出处:
ISSN:

关键词: Unenhanced MR angiography multiple inversion pulses blood suppression inversion time blood flow vascular length phantom

摘要:
Background: This study sought to explore the functional relationship between displayed vascular length and blood suppression inversion time (BSP TI) and flow velocity in a phantom, and to provide a theoretical basis for quantitatively assessing vascular hemodynamic responses using unenhanced magnetic resonance angiography (MRA) and spatial labeling with multiple inversion pulses sequence (SLEEK). Methods: A polyethylene catheter was laid in a long rectangular container filled with pork fat. The entrance of the catheter into the container was connected to a high-pressure syringe filled with normal saline. The high-pressure injector flow rates were set at 0.0, 0.2, 0.4, 0.8, 1.2, 1.6, 2.0, and 2.4 mL/s. SLEEK was performed 19 times for each flow rate with parameter BSP TI values of 50, 75, 100, 150, 200, 300, 400, 500, 600, 700, 800, 900, 1,000, 1,100, 1,200, 1,300, 1,400, 1,600, and 1,800 ms. Maximum intensity projection was employed to reconstruct all SLEEK original images to determine the measurements of the displayed vascular lengths. A regression analysis was undertaken to assess the relationship between the displayed vascular lengths and BSP TI values for each flow rate, and to assess the relationship between the displayed vascular lengths and flow rates at each BSP TI. Results: The displayed vascular length had a linear relationship with BSP TI for each flow rate (P<0.05) (R2=0.754, 0.941, 0.988, 0.988, 0.977, 0.966, and 0.982 for flow rates of 0.0, 0.2, 0.4, 0.8, 1.2, 1.6, and 2.0 mL/s, respectively). The displayed vascular length also had a linear relationship with flow rate for each BSP TI value (P<0.05) (R2=0.914, 0.912, 0.834, 0.989, 0.980, 0.996, 0.992, 0.960, 0.975, 0.979, 0.982, 0.981, 0.976, and 0.993 for BSP TI 50, 75, 100, 150, 200, 300, 400, 500, 600, 700, 800, 900, 1,000, and 1,100 ms, respectively). No significant linear relationship was found between displayed vascular length and flow rate when the BSP TI value was 1,200 ms (P>0.05). Conclusions: Vascular displayed length has a linear relationship to BSP TI for flow ranges from 0.0 to 2.0 mL/s. Vascular displayed length has a linear relationship to flow rate for BSP TI values of 50 to 1,100 ms. Flow rate can be assessed in relation to vascular displayed length.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2020]版:
大类 | 2 区 医学
小类 | 2 区 核医学
最新[2025]版:
大类 | 3 区 医学
小类 | 3 区 核医学
JCR分区:
出版当年[2019]版:
Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
最新[2024]版:
Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING

影响因子: 最新[2024版] 最新五年平均 出版当年[2019版] 出版当年五年平均 出版前一年[2018版] 出版后一年[2020版]

第一作者:
第一作者单位: [1]Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Dept Radiol, Wuhan, Peoples R China
通讯作者:
通讯机构: [1]Huazhong Univ Sci & Technol, Tongji Hosp, Tongji Med Coll, Dept Radiol, Wuhan, Peoples R China [*1]1095 Jiefang Ave, Wuhan 430030, Peoples R China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:622 今日访问量:0 总访问量:452 更新日期:2025-07-01 建议使用谷歌、火狐浏览器 常见问题

版权所有:重庆聚合科技有限公司 渝ICP备12007440号-3 地址:重庆市两江新区泰山大道西段8号坤恩国际商务中心16层(401121)