Suctioning flexible ureteroscopy with automatic control of renal pelvic pressure: a porcine model

| 认领 | 导出 |

文献详情

资源类型:
WOS体系:

收录情况: ◇ SCIE

单位: [1]Nanchang Univ, Dept Urol, Affiliated Ganzhou Hosp, 17 Hongqi Ave, Ganzhou 341000, Jiangxi, Peoples R China [2]Dongyang Hosp Guangfeng City, Guangfeng 334600, Jiangxi, Peoples R China [3]Dermatol Inst Gan Cty, Ganzhou 341100, Jiangxi, Peoples R China [4]Second Peoples Hosp Nankang Dist, Ganzhou 341400, Jiangxi, Peoples R China [5]Chinese Tradit Med Hosp Pinghu City, Dept Urol, Ping Hu 314200, Zhejiang, Peoples R China [6]Huazhong Univ Sci & Technol,Tongji Hosp,Dept Urol,Tongji Med Coll,Wuhan 430030,Hubei,Peoples R China
出处:
ISSN:

关键词: Intelligent control pressure feedback control irrigation and suctioning platform suctioning sheath of flexible ureteroscope renal pelvic pressure

摘要:
Objective: We designed a novel technique of suctioning flexible ureteroscopy with automatic control of renal pelvic pressure (RPP) by a patented system including a pressure-measuring ureteral access sheath (UAS) and an irrigation and suctioning platform. We sought to compare RPP values measured by the patented system and a nephrostomy catheter at different phases of flexible ureteroscopy. Materials and methods: Eight pigs with a total of 16 urinary tract systems were included. Via a subcostal incision a 6-F catheter was placed into the renal pelvis for RPP measurement by connecting to a pressure monitor. We then introduced the patented UAS retrogradely into the renal pelvis, through which the pressure was measured by the platform. RPP was measured at baseline period, irrigation and suctioning period, and therapeutic period. Results: Twelve renal pelves were successfully established models for pressure-measuring. Baseline RPP was 26.9+/-3.8 mmHg in the platform group and 26.3+/-5.2 mmHg in the nephrostomy group. There was no significant difference on RPP between the 2 methods either at irrigation and suctioning period (-5.21+/-2.11 vs. -3.59+/-1.45) or therapeutic period (-5.81+/-2.47 vs. -3.73+/-2.19). Conclusion: Renal pelvic pressure can be accurately and effectively monitored and controlled by our patented system.

基金:
语种:
被引次数:
WOS:
中科院(CAS)分区:
出版当年[2015]版:
大类 | 4 区 医学
小类 | 4 区 医学:研究与实验
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 医学:研究与实验
JCR分区:
出版当年[2014]版:
Q4 MEDICINE, RESEARCH & EXPERIMENTAL
最新[2023]版:
Q4 MEDICINE, RESEARCH & EXPERIMENTAL

影响因子: 最新[2023版] 最新五年平均 出版当年[2014版] 出版当年五年平均 出版前一年[2013版] 出版后一年[2015版]

第一作者:
第一作者单位: [1]Nanchang Univ, Dept Urol, Affiliated Ganzhou Hosp, 17 Hongqi Ave, Ganzhou 341000, Jiangxi, Peoples R China
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:434 今日访问量:0 总访问量:419 更新日期:2025-05-01 建议使用谷歌、火狐浏览器 常见问题

版权所有:重庆聚合科技有限公司 渝ICP备12007440号-3 地址:重庆市两江新区泰山大道西段8号坤恩国际商务中心16层(401121)